KI-DISCUS Servicing and Calibration
Containment Technology Ltd offer KI-DISCUS servicing and calibration. The service is carried out at our UK manufacturing facility in Dorset.
Specification for service and calibration of KI-DISCUS
-
On receipt of KI-DISCUS unit, carry out our functional test, inspect unit and submit report of work/parts recommended that are not covered within the KI-DISCUS calibration service. Carry out such work as agreed.
-
Calibration service to include: a) Check integrity of internal wiring and external cablesand associated connectors. b) Replace metallised spinning disc. c) Examine bore and condition of nebuliser nozzle and delivery rate of peristaltic pump. d) Check spray generator bearing assembly for wear. e) Check condition of and tension on drive belt. f) Clean opto sensor and check distance. g) Check condition of cones in sampling heads. h) Check vacuum at sampling heads and verify accuracy of vacuum guage in tool kit. i) Examine and replace as necessary all 'O' rings. j) Calibrate speed controller to spray generator to achieve 28000 rpm +/- 500 rpm. k) Issue Calibration Certificate
-
Arrangements can be made to collect and re-deliver the unit by a specified carrier conversant with handling KI-DISCUS equipment. |